Objective: The purpose of this study was to separate the enantiomers of amlodipine by High Performance Liquid Chromatography (HPLC) using ovomucoid (OVM) as chiral selector, respectively by Capillary Electrophoresis (CE) using cyclodextrines and to evaluate the analytical performance of the both proposed methods.
Material and methods: HPLC enantioseparation of amlodipine was performed on an HPLC Agilent Technologies 1100 series using as chiral stationary phase an Ultron ES OVM, 150×4.6 mm column with ovomucoid as chiral selector. The stereoselective CE analysis of amlodipine was achieved on Agilent Technologies 7100 CE using uncoated fused-silica capillaries 48 cm x 50 mm and different type of cyclodextrins as chiral selectors.
Results: A mobile phase consisting of 80% Na2HPO4 10 mM at a pH level of 5.0 and 20% ACN, isocratic elution at a flow of 1 ml/min turned to be the optimal experimental conditions for HPLC analysis (R=5.51; α=1.71) with retention times shorter than 10 minutes for the two isomers, tR (S-AML) = 4.63 (min); tR (R-AML) = 5.54 (min). The migration times for amlodipine enantiomers were tm (S-AML) = 8.15 (min) and tm (R-AML)= 8.45 (min) and the optimum CE conditions have proven to be a buffer solution containing 25 mM H3PO4 at pH 3.0 and 20 mM α-CD as chiral selector and a capillary temperature set at 15°C (R=1.51; α=1.03).
Conclusion: The analytical performances of the chromatographic method using OVM as chiral selector are superior to the electrophoretic analysis method but the CE method is more economical and may represent an alternative to the HPLC chromatographic separation.
Tag Archives: CE
The Influence of Some Parameters on Chiral Separation of Ibuprofen by High-Performance Liquid Chromatography and Capillary Electrophoresis
Objective: The aim of the study was to compare the influence of mobile phase composition and temperature on chiral separation of racemic ibuprofen by capillary electrophoresis and high performance liquid chromatography with UV detection.
Materials and methods: Racemic ibuprofen was analysed on a chiral OVM column with an HPLC system 1100 Agilent Technologies, under isocratic elution, by using potassium dihydrogen phosphate 20 mM and ethanol in mobile phase. The flow rate was set at 1 mL/min, UV detector at 220 nm and different column temperatures were tested. For electrophoresis separation an Agilent CE G1600AX Capillary Electrophoresis System system, with UV detection, was used. The electrophoresis analysis was performed at different pH values and temperatures, with phosphate buffer 25 mM and methyl-β-cyclodextrin as chiral selector.
Results: The chromatograhic analysis reveals a high influence of mobile phase pH on ibuprofen enantiomers separation. An elution with a mixture of potassium dihydrogen phosphate 20 mM pH=3 and ethanol, at 25ºC, allowed enantiomers separation with good resolution in less than 8 min.
Conclusions: The proposed HPLC method proved suitable for the separation of ibuprofen enantiomers with a good resolution, but the capillary electrophoresis tested parameters did not allow chiral discrimination.